AutoRAE Lite User’s Guide
10
Fail:
A red LED indicates the bump test or calibration failed.
If a bump test or calibration fails, check the calibration gas pressure and try again. If it fails again, it is
possible the sensor is depleted and must be replaced.
7.1 Status LED Explanation
LED
Red
Green
Solid
Orange
Blinking Orange
Power
Battery low, no external
power
External
power or
battery OK
N/A
N/A
Bump
Fail
Pass
N/A
Processing
Cal.
Fail
Pass
N/A
Processing
Service
Request
No SD card, or
communication failure, or
SD card space is very low
N/A
N/A
SD card is locked or needs
to be formatted, or SD
card space is very low
Whenever a bump test or calibration is performed, the buzzer and LEDs on the ToxiRAE 3 are tested.
If the instrument passes the tests for the buzzer and LEDs, then there is no special indication.
However, if either the buzzer or LEDs test fails, the Bump LED glows green and the Service LED
glows red.