Bump Testing 

You can bump test all or individually selected instruments that are seated in the system’s cradles. 

From the main screen, enter Bump Test by pressing [Y/+] twice (“Function” and then “Select”). 

Note:

 Any cradle that does not have an instrument in it is shown with a gray checkbox. You cannot check/uncheck 

this box. 

To select all instruments in cradles to bump test: 

1.

 

Press [Y/+]. When you check “Bump All,” the “Bump All” checkbox and all of the other checkboxes for 
cradled instruments are automatically checked. 

2.

 

Press [MODE] for “Done.” 

3.

 

Press [Y/+] to start. 

To select specific instruments to bump test: 

1.

 

Press [N/-] to scroll down the list to an instrument you want to select for bump testing. 

2.

 

Press [Y/+] to toggle the selection between unchecked and checked. 

3.

 

Press [MODE] for “Done.” 

4.

 

Press [N/-] to select “Exit,” and then press [Y/+] to exit. 
 

 

 

 

 

 

 

If you did not press [Y/+] to start the process, then a countdown begins. When the countdown reaches zero, then  
the instruments are bump tested. 

To abort a bump test:

 Press [N/-]. You will be asked if you want to abort the process. Press [Y/+] for “Yes” and  

[N/-] for “No.” 

 
IMPORTANT! 

If an instrument does not pass a bump test, a full calibration is automatically started. 
 

 

 

 