Calibration 

You can calibrate all or individually selected instruments that are seated in the system’s cradles. 

From the main screen, enter Calibrate by pressing [Y/+] (“Function”), and then [N/-] until “Calibrate” is highlighted. 
Then press [Y/+] to select it. 

Note:

 Any cradle that does not have an instrument in it is shown with a gray checkbox. You cannot check/uncheck 

this box. 

To select all instruments to calibrate: 

1.

 

Press [Y/+]. When you check “Calibrate All,” the “Calibrate All” checkbox and all of the other checkboxes 
for cradled instruments are checked. 

2.

 

Press [MODE] for “Done.” 

3.

 

Press [Y/+] to start. 

To select specific instruments to calibrate: 

1.

 

Press [N/-] to scroll down the list to an instrument you want to select for calibration. 

2.

 

Press [Y/+] to toggle the selection between unchecked and checked. 

3.

 

Press [MODE] for “Done.” 

4.

 

Press [N/-] to select “Exit,” and then press [Y/+] to exit. 
 

 

 

 

 

 

 

If you did not press [Y/+] to start the process, then a countdown begins. When the countdown reaches zero, then the 
instruments are calibrated. 

To abort calibration:

 Press [N/-]. You will be asked if you want to abort the process. Press [Y/+] for “Yes” and  

[N/-] for “No.” 

 

 

 

 