147
32.4.13. Switch off continuous transmission in current unit
Syntax:
CU0 CR LF
Possible answers:
CU0_I CR LF
- command comprehended but cannot be executed
CU0_A CR LF
- command comprehended and executed
32.4.14. Set lower threshold
Syntax:
DH_XXXXX CR LF
, where:
XXXXX
– mass format
Possible answers:
DH_OK CR LF
- command executed
ES CR LF
- command not comprehended (wrong mass format)
32.4.15. Set upper threshold
Syntax:
UH_XXXXX CR LF
, where:
XXXXX
– mass format
Possible answers:
UH_OK CR LF
- command executed
ES CR LF
- command not comprehended (wrong mass format)
32.4.16. Read lower threshold
Syntax:
ODH CR LF
Possible answers:
DH_MASA CR LF
- command executed
Frame format:
1 2
3
4-12
13 14 15 16 17
18 19
D H space mass space
unit
space CR LF
Mass
- 9 characters justified to the right
Unit
- 3 characters justified to the left
Summary of Contents for WLY/KTP Series
Page 2: ...2 DECEMBER 2011...