- 118 -
34.1.1. Module technical specification
Maximal number of convertor units
8 388 608
OIML class
III
Number of verification units
6000e
Maximal signal gain
19mV
Maximal voltage per verification unit
3,3 µV
Minimal voltage per verification unit
1µV
Minimal strain gauge impedance
90
Ω
Maximal strain gauge impedance
1200
Ω
Strain gauge converter supply voltage
5V
Strain gauge converters connection
4 or 6 wires + shield
34.1.2. Weighing platform signal wires colours
Strain gauge
sensor
determiners
Colours
according to the
RADWAG
standard
W
eighing module
determiners
+INPUT brown +5V
-INPUT green AGND
+OUTPUT yellow
+IN
- OUTPUT
white
-IN
+SENSE gray
+REF
- SENSE
pink
- REF
SCREEN yellow-green
(according to the rule
of connecting shields)