54
17.3. Start Mass Determination
If the weighing device does not require adjustment or you do not have suitable
amount of adjustment weights, you can only determine start mass.
Procedure:
•
Enter
<Adjustment / Start mass determination>
submenu. Message
<Remove weight>
is displayed.
•
Unload the weighing pan and press
key, message
<Start mass
determination; Please wait...>
is displayed.
•
Upon completion of start mass determination procedure, the scale
displays the
<Adjustment>
submenu.
17.4. Adjustment Report
Adjustment report is automatically generated at the end of each adjustment
process, next it is sent to port selected for
<Peripherals / Printer>
. To declare
report content go to
<Printouts / Adjustment report>
submenu.
For instruction on how to declare adjustment report settings read section
'Printouts'.
Report example:
----------Adjustment report----------
Adjustment type External
Operator Nowak Jan
Date 2018.04.10
Time 13:22:28
Scale S/N 123456
Adjustment result difference 0.0g
-------------------------------------
Signature
.....................................
18. SCALE DATA
Scale data menu provides information on the weighing device and its program.
The parameters serve informative purposes: scale S/N, scale type, software
version, product code, settings printout.
In order to send weighing device settings (all parameters) to printer port, select
<Settings printout>
parameter.
Summary of Contents for PUE HX5.EX
Page 1: ...PUE HX5 EX WEIGHING INDICATOR SOFTWARE MANUAL ITKP 01 04 12 18 EN...
Page 2: ...2 DECEMBER 2018...
Page 96: ...96...