- 56 -
18.1.1. Technical specification of A/D converter
Maximal number of divisions from A/D
converter
8 388 608
OIML class
III
Number of verification intervals
6000e
Max increase of signal
19.5mV
Max voltage per 1 verification interval
3.25µV
Min voltage per 1 verification interval
0.4µV
Min impedance of strain gauge load
cells
80 Ω
Max impedance of strain gauge load
cells
1200 Ω
Strain gauge load cells feeding voltage
5V
Strain gauge load cells type
4 or 6 screen
18.1.2. Strain gauge load cells connection
•
6-wire load cells
For 6-wire strain gauge load cells do the connection as follows:
Fig.40 6-wire strain gauge load cell connection