- 40 -
11.1. Autozero function
For ensuring accurate balance readaouts „AUTOZERO” function has been
introduced. This function automatically controls and corrects zero readouts.
When this function is active several measuring results are compared at
regular intervals. If value differentiating these results is lower than declared
AUTOZERO range, e.g. 1 reading unit, than zeroing is automatic. Markers of
stable result -
and zero readout -
are displayed.
When AUTOZERO function is active each measurements start with an exact
zero which is useful. Although there are cases when this function can be a
drawback. One of such examples might be putting load on a pan very slowly
(e.g. pouring the load). Zero readout correction system in such a case might
distort readout of a real mass of the load.
Procedure:
•
Enter user parameters window
,
•
Select or unselect
<Autozeroing>
parameter
- autozeroing function on
- autozeroing function off
11.2. Median filter
Median filter eliminates short-term impulse noise (e.g. mechanical shock).