- 75 -
-
enter <P4 DEVICES> menu
-
enter menu group: <P4.1 COMPUTER>
-
select port to which a computer is to be connected,
COM 1
or
COM 2
– RS 232 port, for connecting a computer
USB 2
type B – USB port, for connecting a computer
USB
FREE LINK
- USB port, type B, for connecting a computer. Tool for entering data, it
functions as a keyboard. Upon proper modification of non-standard printout and sending
proper command from the computer, or upon pressing ENTER key located on the
operation panel, the data from non-standard printout is entered directly to computer
programs, e.g. Excel, Word, Notepad etc.
Select <USB FREE LINK> port and configure its settings in order to enable correct
cooperation of the balance with computer programs:
SPREADSHEET – value ‘YES --.—‘ – cooperation with Excel type program, transferred
decimal values are separated by dot (.)
– value ‘YES --,--‘ – cooperation with Excel type program, transferred
decimal values are separated by coma (,)
– value ‘NO’ – cooperation with other programs (mass value
transferred in a form of text),
DELAY – if the transferred information is incomplete, i.e. part of it gets lost (due to auto-
entering or auto-formatting of data by computer program) during cooperation with Excel
type program, set the parameter value to <2>. If the transferred information is still
incomplete, the parameter value must be higher. The value ranges between 0 (quick data
transfer) and 9 (slow data transfer – around 10 characters per second).
WIFI
– WIFI port
13.1.2. Continuous Transmission
Continuous transmission parameter enables turning on/off continuous transmission of
a measurement result.
Available options:
•
Continuous transmission in a calibration unit <CAL. UNIT>, independently from currently
chosen weight unit, measurements are transmitted in unit set on a main measuring device
(calibration unit)
•
Continuous transmission in a current unit <CURRENT UNIT> measurements are transmitted in
a current unit; the unit changes dynamically whenever the user presses Units button in order to
modify the unit on a display.
•
Continuous transmission off <NONE>
Procedure:
CAUTION!
Continuous transmission may be turned on/off by means of a command sent from a computer (go to
COMMUNICATION PROTOCOL section)
Summary of Contents for PS 10100 R2
Page 95: ...95...