- 87 -
Response format:
1
2
3
4-12
13
14
15
16
17
18
19
U
H
Space
mass
space
unit
space
CR
LF
Mass
- 9 characters with right justification
Unit
- 3 characters with left justification
Set mass value of a single item (only for PARTS COUNTING)
Format:
SM_XXXXX CR LF
, where:
_ -
space
,
XXXXX
– mass format
Response options:
SM_OK CR LF
- command carried out
SM_I CR LF
- command understood but not accessible at this moment (e.g. mode other than
PARTS COUNTING)
ES CR LF
- command not recognised (mass format incorrect)
Set target mass value (e.g. for DOSING)
Format:
TV_XXXXX CR LF
, where:
_ -
space
,
XXXXX
- mass format
Response options:
TV_OK CR LF
- command carried out
TV_I CR LF
- command understood but not accessible at this moment (e.g. mode other than
DOSING)
ES CR LF
- command not recognised (mass format incorrect)
Set reference mass value (e.g. for DEVIATIONS)
Format:
RM_XXXXX CR LF
, where:
_ -
space
,
XXXXX
– mass format
Response options:
RM_OK CR LF
- command carried out
RM_I CR LF
- command understood but not accessible at this moment (e.g. mode other than
DEVIATIONS)
ES CR LF
- command not recognised (mass format incorrect)
Value release
Format:
SS CR LF
Response options:
SS_OK CR LF
- command understood and in progress
Command’s function is similar to function of PRINT button located on an overlay,
Internal adjustment
Format: IC CR LF
Response options:
IC_A CR LF
- command understood and in progress
IC_D CR LF
- calibration completed
IC_A CR LF
- command understood and in progress
IC_E CR LF
- time limit exceeded while waiting for a stable measurement result
IC_I CR LF
- command understood but not accessible at this moment
Disable automatic internal adjustment of the balance
Format: IC1 CR LF
Response options:
IC1_I CR LF
- command understood but not accessible at this moment
IC1_E CR LF
- operation disabled, e.g. for verified balance
IC1_OK CR LF
- command carried out
Operation is disabled for verified balances.
For non-verified balances the command inhibits internal calibration until it is enabled via IC0
command or until the balance is turned off. The command does not modify settings specifying
calibration start.
Enable automatic internal adjustment of the balance
Format: IC0 CR LF
Response options:
IC0_I CR LF
- command understood but not accessible at this moment
Summary of Contents for PS 10100 R2