![RADWAG AS R2.PLUS User Manual Download Page 93](http://html1.mh-extra.com/html/radwag/as-r2-plus/as-r2-plus_user-manual_3660555093.webp)
- 93 -
Response:
EVG_0_OK<CR><LF>
- currently set ambient conditions unstable
Set filter
Format: FIS_n <CR><LF>
Response options:
FIS_OK <CR><LF>
– command carried out
FIS_E <CR><LF>
– error in-course of command execution, no parameter or incorrect
format
FIS_I <CR><LF>
– command understood but not accessible at this moment
n
– parameter, decimal value determining filter number
n →
1 – very fast
2 – fast
3 – average
4 – slow
5 – very slow
CAUTION!
The numbering is assigned to a particular filter name and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, filter settings are
assigned to the working mode.
An example:
Command:
FIS_3<CR><LF>
– set average filter
Response:
FIS_OK<CR><LF>
– average filter set
Give current filter
Command overview:
On receipt of a command, the terminal sends information on currently set filter.
Format:
FIG <CR><LF>
Response options:
FIG_x_OK<CR><LF>
- command carried out, returns set filter
FIG_I <CR><LF>
- command understood but not accessible at this moment
x
- parameter, filter symbols
Example:
Command:
FIG<CR><LF>
-
give current
filter
Response:
FIG_2_OK<CR><LF>
- currently set
average filter
Set Value release
Format: ARS_n <CR><LF>
Response options:
ARS_OK <CR><LF>
– command carried out
ARS_E <CR><LF>
– error in-course of command execution, no parameter or
incorrect format
ARS_I <CR><LF>
– command understood but not accessible at this moment
n
– parameter, decimal value determining value release options
n →
1 – fast
2 – fast+reliable
3 – reliable
CAUTION!
The numbering is assigned to a particular value release option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, value release settings
are assigned to the working mode
An example:
Command:
ARS_2<CR><LF>
– set value release parameter to
fast+reliable option
Response:
ARS_OK<CR><LF>
– fast+reliable option set
Summary of Contents for AS R2.PLUS
Page 6: ...6 1 GENERAL INFORMATION 1 1 DIMENSIONS AS PLUS series PS 200 2000 PS 1000 PS 2100 PS 10100...
Page 97: ...97...
Page 98: ...98...