5.2
After first-time use
If AED was operated for an actual emergency situation, AED maintenance manager
shall check-up followings to make AED ready to be used again anytime for everyone.
5.2.1
Exterior inspection after AED operation
Check exterior appearance of AED, and if it is necessary, please remove
contamination. For damage recovery on AED, please contact customer
technical support
5.2.2
Components management after AED operation
If AED pads were used or unpacked, please replace the pads to new ones
If AED has delivered electric shocks at least over 3 times, and life time has
expired (2 years from manufacturing date), RADIAN suggests replacing
the battery to a new one. Suggested battery life is 4 years on standby mode
6
Troubleshooting
6.1
Self-diagnosis
Power-On Self-test
(Operation mode)
Every time when power button is pressed
Flash ROM test
Charge/Discharge test
Battery Output level test
Regular Self-test
(Stand-by mode)
Automatically performed once per month on 1
st
day
Flash ROM test
Charge/Discharge test
Battery Output level test
If a problem is found in the product during the self-diagnosis test, status indicator
shows X and when turned on, error audio guidance and beeping sound will be
present. In this state, the product will not operate properly.