Temperature requirement for stable operation/storage
Battery and AED pads of HR-503 can be stored from 0
℃
to
40
℃
for short term period(under 5 days), but recommends at
least 1 hour of storage at a room temperature condition prior to
actual AED operation.
For long-term storage(over 5 days), RADIAN recommends
maintaining AED with temperature range from 0
℃
to 40
℃
. AED
pads and battery may have significant performance degrade with
the long-term storage(over 5 days) at out of range of 0
℃
~40
℃
temperature condition.
Operator is to be well-informed of method of AED operation and
CPR in advance
Some national standards restrict AED to be sold by or on the order
of a physician or practitioner licensed by law in which operator
practices to use or order the use of the AED
The AED requires periodic maintenance for followings:
•
Battery manufacturing date check
•
AED pads expiration date check
Do not use AED for CPR training, experiment, and demo
HR-503 is only used for sudden cardiac arrest patient.
Any HR-503 operation on a simulator or a mannequin may cause
serious malfunction or damage on AED
Compliance with national and regional regulation
AED maintenance and authorization, responsibility of usage shall
be followed by relevant national and regional regulation
Product disposal
The AED is classified as electronic device. AED shall be disposed
in a way required by the local guideline
HR-503’s CPR guidance is consistent with 2015 guidelines
defined by AHA and ERC