7

 

Technical data 

7.1

 

Physical characteristics 

Item

 

Description

 

Product Type

 

Semi-automatic external defibrillator

 

Dimensions

 

223.4(W) x 311.7(L) x 91.9(H)mm

 

Weight   
(Including battery and pads)

 

2.5kg

 

 

7.2

 

Environmental characteristics 

Item

 

Condition

 

Detail

 

Temperature range

 

Standby 
- Short-term*   
(less than5 days) 
- Long-term 
Operation

 

 
0 ~ 40

  (32 ~ 104

 
0 ~ 40

  (32 ~ 104

)

 

0 ~ 40

  (32 ~ 104

Humidity

 

 

5 ~ 95%   
(non-condensing)

 

Altitude

 

 

0 to 4,575m   
(0 to 15,000ft)

 

* It is recommended for AED and AED pads to be stored at temperature between 

20

~30

  for normal operation. 

 

 

 

 

 

 

 

Summary of Contents for HEART GUARDIAN HR-503 HEART GUARDIAN HR-503 A

Page 1: ...HR 503 User Manual HR 503 User Manual Rev 0 20 10 2017 ...

Page 2: ... every user of the AED can read the manual Contents of the manual may be changed without prior notification or agreement by Radian This manual can be applied to other models Plagiarism quotation distribution of this manual is prohibited without the consent of RADIAN All copyrights in this manual are reserved by RADIAN 1610 Halla Sigma Valley 53 Gasan Digital 2 ro Geumcheon gu Seoul Korea 82 2 852 ...

Page 3: ...cations of use 4 2 Getting started 4 3 Preparation 4 3 1 AED operation 4 3 2 Data management 4 3 3 Guidance for maintenance Chapter 5 AED location and periodical maintenance 5 1 After first time use 5 2 Troubleshooting Chapter 6 Self diagnosis 6 1 AED conditional alarm 6 2 Troubleshooting 6 3 Technical data Chapter 7 Physical parameters 7 1 Environmental parameters 7 2 Functionality 7 3 Applicable...

Page 4: ...that may cause injury or death Caution The symbol implements hazards that may cause an electric shock product damage and malfunction Reference Additional reference data is referred Dangerous Voltage The defibrillator output has high voltage and may present a shock hazard Additional info Additional information is described on manual Type BF Equipment Non Sterile Product Part is not sterilized Non R...

Page 5: ...be replaced after marked date Do not expose to sunlight Do not expose to direct sunlight WEEE Waste Electronic Electrical Equipment Separate collection for waste electrical and electronic equipment Direct Current DC Battery Type Lithium manganese oxide Battery Temperature Limit EC Representative Serial Number LOT Batch code Product Model Number Do not use if package is damaged Single time use only...

Page 6: ...ry Power Button Press the button to operate AED Electric Shock Button Audio and LED guidance will notice to press the shock button when it is needed Adult Child Mode Select Button Select Adult Child mode prior to AED pads attachment Default Adult mode Language Select Button Select a representative language Default Korean Button function may be different as per model HR 503 A Adult Mode Button HR 5...

Page 7: ...t button Be careful of electric shock hazard Electric shock can be delivered to operator or bystander through unexpected ways To avoid this hazard during AED operation keep followings Do not use AED in standing water underwater rain snowing Move patient to a dry area Confirm and instruct anyone to stay away from the patient prior to pressing Shock Button When patient is contacted with any conducti...

Page 8: ...m the patient Touching the patient during the ECG analysis phase can cause interference with the diagnostic process that may cause increased analysis time or result analysis error Avoid contact with the patient while analysis is on progress AED pads are single use only and have expiration date Do not reuse AED pads Also always keep the AED package sealed prior to the actual use If the sealed packa...

Page 9: ...ld by or on the order of a physician or practitioner licensed by law in which operator practices to use or order the use of the AED The AED requires periodic maintenance for followings Battery manufacturing date check AED pads expiration date check Do not use AED for CPR training experiment and demo HR 503 is only used for sudden cardiac arrest patient Any HR 503 operation on a simulator or a mann...

Page 10: ...2 Defect caused by unauthorized repair or modification 3 2 3 Defect caused by external shock including drop 3 2 4 Defect caused by natural disaster such as fire earthquake flood and thunder stroke 3 2 5 Defect caused by use storage in extreme temperature or humidity environment please refer to Chapter 5 Guidance for maintenance 3 2 6 Unauthorized alternation of purchase date purchaser name vender ...

Page 11: ...the defect Once Radian confirm the defect is caused by one of the manufacturing development or delivery procedure repair exchange may be provided without any shipping charge Shipping will be completed within maximum of 50 days from the day the product arrives at RADIAN Corp which are automatically added up to remaining warranty period ...

Page 12: ...ng training schedule 4 1 Product overview HR 503 a portable electric device operated by battery power is an easy to use Semi Automated External Defibrillator The AED automatically analysis patient s ECG and identifies and advise electric shock if it is necessary Nearly every procedure for AED operation is guided through voice message and beep sound as well as the visual guidance with picture descr...

Page 13: ...s may be different as per model refer to p20 p21 ⓔ AED Pad Connector For connection of AED pads and the device ⓕ Open Button For opening AED cover ⓖ Handle For carrying AED safely ⓗ Procedure LED Guidance LED light informs CPR and AED operating procedure to user ⓘ AED Pads For AED to analyze ECG and deliver shocks ⓙ Cover Prevents external substances exposure to AED ⓚ Battery Indicator For indicat...

Page 14: ...e select button Battery Type BT 503 LiMnO2 12V DC 4 2A Number of shock At least 200 times with new battery Child mode AED Pads Type P 503 AED electric pads single use 2 lead wire 1 8m 2 hole plug Product Size mm 223 4 W x 311 7 L x 91 9 H mm Product Weight kg 2 1kg battery AED case and electric pads excluded Product Color Orange Basic Components AED device AED portable case HR B3 AED pads P 503 Ba...

Page 15: ...le for both adult child Size 150 W x 190 D x 15 T mm Weight 100g Warranty 2 years in case of damage caused by user s mistake repair exchange will be charged Life time Single use only 2 years for unpacked sealed product from the manufacturing date BT 503 HR 503 Battery Model BT 503 Usage Electric power supply to AED Type LiMnO2 DC 12V 4 2Ah Size 175 W x 60 D x 25 H mm Weight 285g Warranty 4 years i...

Page 16: ...button operation adult mode on patients greater than 8 years old or over 25 kg child mode on patients between 1 and 8 years of age or up to 25 kg Default is set as adult mode when AED is released It is highly recommended for every potential user to complete periodic training for CPR and HR 503 operation Please contact national local representative for Heart Guardian series to find out detailed CPR...

Page 17: ...representative Manual and quick guide is also available at www radian co kr en for downloading I HR 503 AED Device bag included II HR 503 User Manual III P 503 AED Pads IV BT 503 Battery Non rechargeable V Quick Guide Insert a battery pack BT 503 into the battery slot For safe storage always keep AED in the portable case HR B1 Place AED at a highly visible and accessible location Do not remove sea...

Page 18: ...ult child mode Heart Guardian HR 503 may be used for both adult and child If it is hard to define the patient s weight and age quickly proceed further operation procedure with the default mode adult mode Adult Mode Default Above age of 8 or Minimum weight 25kg Child Mode Below age of 7 or Maximum weight 25kg ...

Page 19: ...ght of 25kg When selected child mode indicating lamp will be turned on Language select button right Language indicating lamp will be turned on depending on the selected language Heart Guardian HR 503 A Button Function Adult mode select button left Select if the patient is above age of 8 or minimum weight of 25kg When selected adult mode indicating lamp will be turned on Child mode select button ri...

Page 20: ...ton right Select if the operator is a non expert When selected non expert mode indicating lamp will be turned on Heart Guardian HR 503 C Button Function Adult mode select button left Select if the patient is above age of 8 or minimum weight of 25kg When selected adult mode indicating lamp will be turned on Child mode select button left Select if the patient is below age of 7 or maximum weight of 2...

Page 21: ...may be necessary for better AED pads attachment 4 2 2 4 Attach AED Pads to the patient 1 Take the AED pads out from the cover 2 Remove the package and the covering stickers 3 Attach the pads to patient s chest Saving time is very important for sudden cardiac arrest patient Time shall not be wasted on taking off the clothes Remove patient s top as quickly as possible even by cutting or ripping it ...

Page 22: ...tay away from the patient Press the flashing orange shock button Electric shock is delivered to the patient Electric shock has been delivered After electric shock perform CPR procedure following audio guidance Audio guidance Start CPR Audio guidance Compress the chest 30 times Beeper rings 30 times Audio guidance Start mouth to mouth CPR 2 times There is no audio guidance in the non expert mode in...

Page 23: ...the product operates as below Audio guidance Firmly attach the pads to the patient s bare skin Even though electric energy charge is completed if user does not push blinking shock button for more than 15 seconds AED will conduct discharge the electric energy instead of delivering electric shock to patient through AED pads Audio guidance Shock button was not pressed Be careful of electric shock Ele...

Page 24: ...e to a telephone in order to call emergency aid service immediately in case of an emergency 5 1 2 All components including P 503 electric pads shall be kept together 5 1 3 AED shall always be stored with the battery installed 5 1 4 Periodically check AED operation and expiration date of battery and AED pads 5 1 5 The product shall be stored operated in following environmental conditions Long term ...

Page 25: ...rated properly in emergency situation When status indicator shows X battery pack shall be replaced immediately Periodically check manufacturing date indicated on battery label and replace to new one if it is expired 4 years from manufacturing date Replacing battery pack Pull out the battery pack to be replaced on backside of the AED by pressing a lever on left side of the battery pack Push in the ...

Page 26: ...e use of unofficial battery may cause trouble malfunction please use the official battery provided by RADIAN AED pads maintenance AED pads expiration date shall be checked periodically Expired AED pads shall be replaced to new one immediately Expiration date is marked on the package If sealing package of AED pads is ripped or damaged do not use the pads and replace it to new one When AED pads are ...

Page 27: ...ric shocks at least over 3 times and life time has expired 2 years from manufacturing date RADIAN suggests replacing the battery to a new one Suggested battery life is 4 years on standby mode 6 Troubleshooting 6 1 Self diagnosis Power On Self test Operation mode Every time when power button is pressed Flash ROM test Charge Discharge test Battery Output level test Regular Self test Stand by mode Au...

Page 28: ... state of normal operation Prompts when error is found during operation Prompts the current state of battery Procedure LED guidance Orange Blink Guides the current CPR procedure Shock Button Light Orange Blink Prompts the state of preparation for electric shock ...

Page 29: ...contact the vender for A S Operating Audio message The battery is low please change the battery immediately Replace the battery to a new one Audio message System error Contact the vender for A S Audio message System will be terminated due to no attachment of the pads Attach AED pads to the patient properly Attach AED pads firmly to the skin of the patient Audio message Attach the pads to the patie...

Page 30: ...Including battery and pads 2 5kg 7 2 Environmental characteristics Item Condition Detail Temperature range Standby Short term less than5 days Long term Operation 0 40 32 104 0 40 32 104 0 40 32 104 Humidity 5 95 non condensing Altitude 0 to 4 575m 0 to 15 000ft It is recommended for AED and AED pads to be stored at temperature between 20 30 for normal operation ...

Page 31: ...with a low on battery the charging time can be significantly prolonged Operating time For a new fully charged battery pack at 20 30 C 200 discharges at maximum energy or 10 hours of continuous ECG monitoring Automated Self tests Monthly every 1st day for each month Battery charge discharge internal electronics Audio Guidance Alerts Guidance for AED operation and CPR procedure voice record Beep sou...

Page 32: ...ation Free fall drop Degree of Protection Particle and water ingress IEC 60529 IP54 EMC 7 5 Characteristics of P 503 AED Pads Item Description Type Pre gelled self adhesive disposable non polarized defibrillation pads Shelf life 2 years Disposal Abide by regional national regulations ...

Page 33: ... Rhythm VF IEC 60601 2 4 AHA s recommend 90 Shockable Rhythm VT IEC 60601 2 4 AHA s recommend 75 Non shockable Rhythm NSR IEC 60601 2 4 AHA s recommend 99 Non shockable Asystole IEC 60601 2 4 AHA s recommend 95 Non shockable all other rhythms IEC 60601 2 4 AHA s recommend 95 Rhythms Standard advised by AAMI Minimum Test Sample Size Test Sample Shock Decision No Shock Decision Performance Coarse VF...

Page 34: ... 1610 Halla Sigma Valley 53 Gasan Digital 2 ro Geumcheon gu Seoul Korea 82 2 852 1122 Info radian co kr www radian co kr ...

Reviews: