Warning and Caution
Please refer to the following for safety use of HR-501
Certain contents may be indicated delicately in other part
Do not use if the patient shows any reaction or is
consciousness
HR-501 is a portable defibrillator manufactured for
ventricular fibrillation patients. Prior to AED operation, it is
very important to check patient’s consciousness and
respiration condition; do not use if the patient is conscious
and reacting
The AED is set to “Adult Mode” as default
In adult mode, AED delivers about 150J of electric shock to
patient. In child mode, about 50J of electric shock is set to be
delivered, if patient is under the age of 7 or weighting below
25kg, please set “Child Mode” by moving Adult/child mode
switch to the left and attach the pads firmly to appropriate
spots
Adult
Child
Be careful of electric shock hazard
Electric shock can be delivered to operator or bystander
through unexpected ways. To avoid this hazard during
AED operation, keep followings:
Do not use AED in standing water, underwater, rain,
snowing. Move patient to a dry area
Confirm and instruct anyone to stay away from the
patient prior to pressing ‘Shock Button’
When patient is contacted with any conductive materials
such as metals, stay away from the conductive materials