4.1.2
Specification
Product Type
Semi-automated External Defibrillator
Auto/Manual Diagnosis
Automatic diagnosis (battery power, circuit power,
button operation checks)
ECG complete analysis
FFT ECG Analysis
Step by Step Audio
Guidance
Voice record (English)
Compatible Adult/Child
Adult/child mode switch (default)
Charge Method
Hi-Cap charge
Output Range
150J
±
10% (Adult) / 50J
±
10% (Child)
Data Storage
108 times of ECG log can be recorded
Control Button
Power, Analysis, Shock, Adult/child mode switch
Battery Type (BT-303)
LiMnO
2 (
12V DC 4.2A)
Number of shock
At least 200 times (New battery)
AED Pads Type (P-303)
AED electric prods(single-use), 2 lead wire(2m),
4-hole plug
Product Size (mm)
225(W) x 320(L) x 85(H)mm
Product Weight (kg)
2.2kg (battery included, AED case and electric
pads excluded)
Product Color
Red
Basic Components
AED portable case(HR-B1), AED pads(P-303),
Battery(BT-303), User Manual, Quick guide