5601 User’s Manual
2-4 Installation
Racal Instruments
Ó
2003
1. Locate LK2 near the bottom of the board, as shown in
Figure 2-1.
2. Remove the shorting jumper if you want to float the
input/output grounds from case ground.
3. Leave jumper LK2 in place if you want to connect
input/output ground to case ground.
NOTE
Normal amplifier operation is recommended
with circuit ground connected to case ground
(float disabled). This minimizes susceptibility
to system noise. Enable the floating-ground
configuration only for applications requiring a
floating ground.
Figure 2-1. Float Enable/Disable Jumper (LK2)
Calibration
The recommended calibration interval is three years.
Calibration should be performed by qualified personnel only.