1266 Product Specifications A-3
Appendix A
Monitor
Specifications
Backplane
Specifications
•
Automatic daisy-chain jumpering for BUS GRANT and IACK signals.
•
Full differential distribution of CLK10.
•
ACFAIL* and SYSRESET* in full compliance with the VMEbus and
VXIbus Specifications.
•
Surface mount construction and no sockets for maximum reliability.
Power Specifications
This section describes 1266 mainframe power specifications for the
Standard configurations.
Size:
C-Size
Slots:
4 available
Connectors:
P1/P2
Shared Memory:
n/a
VXI Buses:
per VXIbus Spec, Rev 1.4
Category
Specifications
Interface
Description
Power Switch
Stdby
- AC power is applied to the mainframe, but not to installed VXI instruments.
On
- AC power applied to the mainframe, and DC to installed VXI instruments.
Voltages Indicator
Green
- all voltages on the VXI backplane are present. (Does not necessarily indicate
that voltages are within the VXI specification.)
Flashing Amber
- one or more voltages is missing or incorrect. The backplane
voltages and allowed variations are listed below:
Voltage Allowed Variation
+5V +4.875V to +5.125V
+12V +11.64V to +12.6V
-12V -12.6V to -11.64V
+24V +23.28V to +25.2V
-24V -25.2V to -23.28V
-5.2V -5.46V to -5.044V
-2V -2.1V to -1.9V
Temp Indicator
Green
- the power supply temperature is acceptable.
Flashing Amber
- the power supply is approaching thermal shutdown.
Fan Indicator
Green
- mainframe cooling fan is operating properly.
Flashing Amber
- mainframe cooling fan is not operating.
Fan Speed Switch
Full
- the mainframe cooling fan is operating at full speed.
Var
(iable) - the mainframe cooling fan is providing adequate cooling (default setting).