User Manual 1260-30
Module Specific Syntax 3-1
Chapter 3
MODULE SPECIFIC SYNTAX
1260-30 Module
Specific Syntax
The Module Specific Syntax for the 1260-30 is required in the use
of the OPEN and CLOSE commands. It will also appear in data
output by the Master in response to the PDATAOUT and PSETUP
commands
.
Syntax
The Module Specific Syntax for the 1260-30 Signal
Multiplexer/Scanner module is as follows:
OPEN <module address>.<group number><channel>
where <module address> is the address.
<group number> is the number of switching groups
implemented in the module. The
range of values for <group
number> is up to 0-7
<channel> is the path in the group to be switched.
For 1260-30 configurations containing
more than one group, the values for
<channel> are repeated for each group.
NOTE
The <module address> used here is NOT the VXIbus
defined logical address of the 1260 Master. It is
peculiar to the 1260 Series and describes the
switching module in relation to the Master. This
address corresponds to the binary value of the switch
setting of SW1 on the switching module PCB.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com