1260-132 User Manual
Module Operation 3-4
OPEN (@8(5))
Open channel 5.
OPEN (@8(10))
Open channel 10.
CLOSE (@8(0,9))
Close channel 9 on the 1260-132.
(Note that Channel 0 is used as
the common output and must be
closed to complete the path)
CLOSE (@8(0,23))
Close channel 23 on the 1260-
132.
OPEN (@8(0:23))
Open channels 0 through 23 (all
channels) on the 1260-132.
CLOSE (@8(0,10:22))
Close channels 0, 10, through 22
on the 1260-132.
Power Up Relay
Default State
Normally, the default state for all relays for the 1260-132 is to be
open on power up. The 1260-01T controller or the 1256 will set this
state by default . This may or may not be what the user desires.
In the event that the user would like to change the state of the
relays after power up it can be accomplished by doing the following:
1)
Close each of channels on the 1260-132 that you would like
to configure as closed for the power up state. For example, if the
module address = 3:
CLOSE (@3(0,1))
3)
Repeat step #2 for EACH 1260-132 in the system
4)
Execute the "*SAV 0" command to save the present states
of all relays into non-vol.
After this, whenever the Option -01T or 1256 is powered on, the
specified relays will be closed (until specifically commanded to be
open).