1260-00C User Manual
Local Command Set 4-41
where
<byte value>
is the new value of the ESE
register.
Action: Sets ESE to
<byte value>.
Example:
Set the ESE register to 45.
*ESE 45
*ESE?
488.2
Intent: Get the contents of the ESE register.
Query
Syntax:
*ESE?
Response:
<ESE val>
<CRLF>
where
<ESE val>
is the current value of the ESE
register. The default value is FFh.
*ESR?
488.2
Intent:
Read and clear the Standard Event Status register
(ESR).
Query
Syntax:
*ESR?
Response:
<ESR val>
<CRLF>
<ESR val>
is the current value of the ESR.
*IDN?
488.2
Intent:
Get the 1260-00C's manufacturer, model, serial
number, and firmware level.
Query
Syntax:
*IDN?
Response:
"Racal-
Dana","1260-00C",<serial
number>,<firmware version>
<CRLF>
*OPC
488.2