1260-00C User Manual
Local Command Set 4-24
Error sending Logical Address <la> Word
Serial command <hex cmd
val><CRLF><space><space><ws error>
<CRLF>
if the command failed.
<la>
is the logical address of the device to which
the broadcast failed.
<cmd val>
is the value of the Word Serial
command, in decimal.
<hex cmd val>
is the
value in hexadecimal.
For Word Serial queries,
<ws response>
is the
Word Serial response of the device at Logical
Address
<la>
. For Word Serial commands
<ws
response>
is 0.
<Word Serial command name>
is the name of
the command name as shown in the previous table.
<ws error code>
is a decimal value that can be
interpreted by converting it to a binary bit pattern. A
value of 1 in the bit positions shown in the following
table indicates that an error occurred during the
attempt to broadcast the Word Serial command:
Bit
Word Serial Error
0
Word Serial Command completed successfully (no
Word Serial error)
1
Timeout waiting to send Word Serial command to
device at <
la
>
2
Timeout waiting for Word Serial response from
device at <
la
>
3
Device at <
la
> did not recognize the command
6
Multiple query error
10
Read Protocol error not supported
13
Read Ready (RR) violation
14
Write Ready (WR) violation
None of the other bits has significance in this
context.
<ws error>
is a string explaining the Word Serial
error as shown in the previous table.
Example:
Broadcast the Identify Commander Word Serial