QwikCheck™
Sperm Quality Analyzer Guide Version 1.00 I-Button WHO 5th 06_AUG_2012
31
APPENDIX 10: Service Report
SQA SERVICE SUPPORT
Parameter Report
Device number: _______ Software Version: ___________ Date: __________
Instruct the user to run a SERVICE report. For the QwikCheck GOLD system go to:
MAIN MENU > PRINT
SELF TEST DATA AND DEFAULT SETTINGS>SELF TEST DATA.
Calibration parameters:
Fill-in the USER REPORT column with the calibration parameters found in the INTERNAL DATA SECTION of the
SERVICE DATA REPORT of the QwikCheck GOLD. Contact your local distributor for the initial calibration
parameters. These parameters should not have changed.
Parameter
Service
Report
Item #
User Report
Initial
Calibration
settings
Comments
CONTR.REF1
#1
OD AMPLIF.
#13
MSC AMPLIF
#8
OD VALUE
#15
OD CORR
#16
LB OD AMP
#18
CONTR. Z.L*
#11
*CONTR. Z.L. can be adjusted in the field by a MES trained service technician
Algorithm parameters
Fill-in the User Report values for the following algorithm parameters found in the SERVICE DATA REPORT.
The QwikCheck algorithm settings are defined and should not have changed.
Parameter
Service
Report
Item #
User
Report
Initial
Settings
Comments
MIN.SP.HEIGHT
#2
MIN.SP.WIDTH
#9
MAX.SP.WIDTH
#3
NOISE THRESH
#10
SMI THRESH
#4