QM1014 LO Synthesizer Distribution Unit
User Manual
5.7 SYSTem:SAVEstate [1-5]
This command saves the current setup to non-volatile memory. There are 5 re-writeable memory loca-
tions, specified by choosing an index between 1 and 5. Any of the 5 states can be restored on startup
when specified with the SYSTem:BOOTstate command. The parameters saved are the
RF Power State,
External Reference Enabled, External Referece Switch Override, Tune Frequency
. For a detailed de-
scription of the parameters saved, see the relevent command descriptions in this manual. NOTE: The
MEM_CLR button on the back panel will rewrite the contents of states 1-5 with the contents of state 0
when depressed, serving as a reset to factory defaults.
Syntax
:
SYST
:
SAVE
Space
1-5
?
There are 5 memory locations, numbered 1 through 5. Memory location 0 contains the factory default
settings and is write-protected.
Example
:SYST:SAVE 3
This command saves the current state to memory location 3
Error Message
If the parameter is not in the recognized format, error
-102, "Syntax error"
occurs.
If the parameter is less than 1 or greater than 5, error
-222, "Data out of range"
occurs.
Quonset Microwave
Revision 1.2.1
48