Page
41
of
48
Shutdown
Turn off the unit by holding the power switch in the rear of the unit. The shutdown is
complete when the screen goes dark.
Maintenance and Cleaning
Maintenance
Battery Power
Battery Intended Use
Sofia 2 includes a rechargeable battery. The battery will automatically charge while plugged
into AC power. Sofia 2 will automatically switch to battery power when AC power is not
available.
Battery Replacement
Sofia 2 is shipped with an internal LiPo rechargeable battery with an expected life of
approximately three years.
Use only a Quidel-supplied Sofia 2 replacement battery.
To optimize instrument performance and charging, it is recommended that the battery
remain installed in the instrument.
Before replacing the battery, ensure that there is no cassette in Sofia 2. Then power down
and unplug Sofia 2. Remove the battery cover on the rear of the instrument. Disconnect the
battery cable and carefully slide out the battery (Figure 30).
Attach the new battery and replace the battery cover. Recycle or dispose of the lithium
battery in accordance with all Federal, State, and Local laws. To avoid fire and explosion
hazards, do not burn or incinerate the battery. Batteries shall not be disposed of with
municipal waste.
Sofia 2 Storage
If the Sofia 2 is intended to be stored and/or not used for an extended period of time (> 3
months), unplug the unit from AC power and remove the battery (Figure 31). When ready
to use the Sofia 2 again, install the battery and follow the Installation, Setup, and Operation
section of this User Manual.
Note:
Sofia 2 must be sent to Quidel if maintenance is required. The User should not
attempt any maintenance except for changing the battery and cleaning the external
surfaces and Cassette drawer.