Page
33
of
48
Ensure that
is selected
,
then press
to import Sofia 2 firmware. Sofia 2 will complete a
self-installation and then power up in the normal startup screen. Once complete, it is safe to
remove the USB flash drive.
Import Assay Methods from USB Flash Drive
Insert USB flash drive with the assay method to install.
Use or in the first field to select
Import
.
In the next field, use or to select
Assay Methods
.
Ensure that
is selected
,
then press
to import Sofia 2 assay methods. Sofia 2 will
complete a self-installation and return to the previous screen. Once complete, it is safe to
remove the USB flash drive.
Import Language Files from USB Flash Drive
Insert USB flash drive with the language file to install.
Use or in the first field to select
Import
.
In the next field, use or to select
Language Files
.
Ensure that
is selected
,
then press
to import Sofia 2 language files. Sofia 2 will
complete a self-installation and return to the previous screen. Once complete, it is safe to
remove the USB flash drive.
Run Test (Patient Specimens)
The Run Test screen can also be accessed by selecting
on most screens.
Enter the User ID
The User ID can be entered using the barcode scanner or manually using the on-screen
keypad.
Select the
User ID
field on the Run Test screen. Use the barcode scanner (Figure 25) to scan the
barcode from a User
’s ID Card or manually enter the data using the on
-screen keypad on Sofia 2.
Summary of Contents for Lyme FIA Sofia 2