Installation, Setup, and Operation
Sofia 2 Setup
◼
Place Sofia 2 on the benchtop. The unit is portable and can be moved to a suitable
location for testing. Ensure the countertop is stable, level, dry, and free of
obstructions. Avoid direct sunlight.
◼
Ensure the bench provides adequate space for Sofia 2 and a
6”
minimum clearance
around the unit. There must be space to allow adequate ventilation and to
open/close the Sofia 2 drawer and access the connection ports on the back of the
unit.
◼
Connect the AC Power Cord to the Power Adapter and ensure the cord is seated into
the Power Adapter.
◼
Plug the Power Adapter with the round plug head into the power port on the back of
Sofia 2. Then plug the AC Power Cord attached to the Power Adapter into an
available electrical outlet.
◼
If using the optional Dymo Label Printer, connect to Sofia 2 by plugging into a USB
port on the back of the instrument (Figure 2).
Power Up
Turn Sofia 2 on using the power switch located on the rear panel (Figure 2).
Figure 2
Battery Compartment
Power Port
RJ45
USB Ports (2)
Power Switch