QM-7 SERIES SERVICE MANUAL
PAGE 7.10
JANUARY 2012
SUNRISE MEDICAL
Section 7
4. Remove the actuator and micro-switch assembly
as one part if the entire assembly is to be replaced
(
fi
gure 7.10.1).
5.
Reverse the previous procedure to install
the actuator and micro-switch assembly. Retest
as necessary.
fig. 7.10.1
Recline Actuator Removal
Note:
In order to remove either the recline or the shear actuator, the ISM box and mount must
be removed. The gas strut must also be unbolted on one end, and swung out of the way.
Complete the following steps in order to remove and replace the recline actuator.
1. Locate and Remove the rear shroud (
fi
gure
7.10.2) by using a
fl
at head screw driver to lift the
small head of the fasteners (F) (
fi
gure 7.10.3)
F
F
fig. 7.10.3
fig. 7.10.2