V I I . O P E R A T I O N
Remote Plus Error Codes
In addition to indicating the current state of battery charge, the battery condition meter can also indicate possible problems
with your power chair’s electrical system. If any of the battery condition meter lights are flashing rapidly, the controller may
be indicating an error. Error codes are displayed as a number of flashing lights. For instance, if the first light is flashing
rapidly, the battery voltage is nearly depleted. The table below identifies the individual error codes, probable causes, and
possible solutions. If you get one of these error codes, contact your Quantum Rehab Specialist.
FLASHING LIGHTS
DIAGNOSIS AND SOLUTION
1
The batteries need charging or there is a bad connection to the batteries. Check the
connections to the batteries. If the connections are good, try charging the batteries.
2
The left motor has a bad connection. Check the left motor connection.
3
The left motor has a short circuit to a battery connection. Contact your Quantum Rehab
Specialist.
4
The right motor has a bad connection. Check the right motor connection.
5
The right motor has a short circuit to a battery connection. Contact your Quantum Rehab
Specialist.
6
The power chair is being inhibited by the battery charger. Unplug the battery charger.
7
A joystick fault is indicated. Make sure that the joystick is in the neutral (centre) position
before turning on the controller.
8
A controller system fault is indicated. Make sure that all connections are secure.
9
The parking brakes have a bad connection. Check the parking brake and motor
connections. Make sure the controller connections are secure.
10
An excessive voltage has been applied to the controller. This is usually caused by a poor
battery connection. Check the battery connections.