V I I . O P E R A T I O N
Actuator Keys and Actuator Lights (For Optional Equipment)
Actuator keys and actuator lights are used for optional equipment such as power elevating leg rests. For specific operation
of the actuator keys and actuator lights, contact your Quantum Rehab Specialist.
Horn Key
The horn key activates the horn.
Locking/Unlocking the VSI
The VSI has a feature that enables you to lock your power chair to prevent unauthorised use.
To lock the VSI:
1. With the VSI switched on, press and hold the on/off key. After 1 second, the VSI should beep.
2. Release the on/off key.
3. Push the joystick to the full forward position until the VSI beeps.
4. Pull the joystick to the full rearward position until the VSI beeps.
5. Release the joystick. There should be a long beep.
6. The VSI is now locked.
To unlock the VSI:
1. Press the on/off key and power on the VSI. The maximum speed/profile indicator should ripple up and down.
2. Push the joystick to the full forward position until the VSI beeps.
3. Pull the joystick to the full rearward position until the VSI beeps.
4. Release the joystick. There should be a long beep.
5. The VSI is now unlocked.
NOTE: If the above procedure fails to either lock or unlock the VSI, contact your Quantum Rehab Specialist.
Off-board Charger/Programming Socket
The off-board charger/programming socket is located on the front of the VSI. If you use an off-board charger, the charger
current should not exceed 8 amps. Contact your Quantum Rehab Specialist for more information.
WARNING! Only chargers with Neutrik NC3MX plugs should be connected to the off-board charger/
programming socket. See your Quantum Rehab Specialist for more information.
NOTE: The socket may also be used for reprogramming the VSI. Contact your Quantum Rehab Specialist for
more information.
Motor Connectors
This connects the VSI to the power chair’s motors and motor brakes.
Controller Connector
This connects the VSI to the power chair’s batteries, motors, and motor brakes.
Battery Connector
This connects the VSI to the power chair’s batteries.
Actuator Connector (Optional)
This connects the VSI to the power seat actuator.