Model QS-500 Tubestand
Chapter 3 Assembly & Installation
22
Leveling Support Arm (Standard Equipment Support Arms)
1. Verify that Support Arm is level as shown in Figure 12. If Support Arm is
not level (
±
1.0
°
), verify that the Floor Track and Tubestand are level.
2.
If Floor Track is level, then level Support Arm/Transverse Arm as follows:
a. Loosen four (4) 3/8-16
x 1” hex head cap
screws securing Sup-
port Arm to Vertical
Trolley (see Figure
13).
b. Adjust set screws as
required to level arm
and collimator.
c. Re-tighten four (4)
3/8-16 x 1” hex head
cap screws.
LEVEL PLACED ON
SUPPORT ARM
Figure 12. Checking Collimator/Tube Arm Extension Level
SET SCREWS
(4X)
3/8-16 X 1”
HEX HEAD
SCREWS AND
LOCK-WASHERS
(4X)
Figure 13. Tube Arm Extension Leveling Screws
Summary of Contents for QS-500 Series
Page 4: ...Model QS 500 Tubestand THIS PAGE INTENTIONALLY LEFT BLANK ii Revision History...
Page 6: ...Model QS 500 Tubestand Table of Contents THIS PAGE INTENTIONALLY LEFT BLANK iv...
Page 12: ...Model QS 500 Tubestand THIS PAGE INTENTIONALLY LEFT BLANK Warranty Information x...
Page 13: ...Chapter INTRODUCTION 1 1...
Page 14: ...2...
Page 17: ...5 Chapter SPECIFICATIONS 2...
Page 18: ...6...
Page 22: ...Model QS 500 Tubestand 10 Chapter 2 Specifications Figure 2 Model QS 500 Tubestand Dimensions...
Page 23: ...11 Chapter ASSEMBLY INSTALLATION 3...
Page 24: ...12...
Page 43: ...Chapter OPERATION 4 31...
Page 44: ...32...
Page 48: ...Model QS 500 Tubestand Chapter 4 Operation 36 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 49: ...Chapter MAINTENANCE 5 37...
Page 50: ...38...