Model QS-500 Tubestand
Vertical Column Angle Adjustment
1. Place a double-bubble level, two feet in length minimum, on the front of the
Vertical Column (see Figure 8).
2. If not level, adjustment can be made by moving the Floor Track in or out
from wall. Ensure the Vertical Column is level throughout the full length of
longitudinal travel.
3. Secure the Floor Track to floor using appropriate fastening devices
(depending on floor construction), each capable of supporting 200 lbs. pull-
load.
19
Chapter 3 Assembly & Installation
Figure 8. Checking the Tubestand Vertical Angle
VERTICAL
COLUMN