10-41
sending remote commands to the device. Since the DUT and the 9730 may reside in
separate locations, this was an important safety measure to incorporate.
Some customers have started using the 9730 device in high throughput production lines
and have requested that this safety feature be removed.
Quantum Composers has added
a bypass feature to accommodate these customers under the condition that usage of
this bypass function is an acceptance agreement by the user that they are knowingly
and willingly bypassing a safety feature for the 9730 and accept any liability
associated with bypassing such feature
.
Bypass Instructions
The safety feature can by bypassed by sending a command after the unit has been
powered on. This command can be sent along with any other initialization commands that
may already being sent on startup. This command is a static command such that when the
unit is powered off, it will return to the default safe operation until the initialization
command is sent again. The commands are as follows:
:SPUL:ARMD 0
(This will disable the feature of disabling the channels and not
automatically re-enabling them).
:SPUL:ARMD 1
(This will restore the safety feature of disabling the channels and
not automatically re-enabling them)
:SPUL:ARMD?
(This is a query command to read the current state of the bypass
command)