10-37
Keyword
System Pulse Commands
Parameter
Comments
:SPULse or
PULSe[0]
Subsystem. Contains commands to control the output
pulse generation. Commands without suffix refer to the
currently selected logical instrument. See INSTrument
subsystem.
:STATe
<boolean value>
Enables / Disables the output for all channels.
Command is the same as pressing the RUN/STOP
button.
:PERiod
<numeric value>
Sets the T
0
period.
:MODe
SINGle / BURSt
Sets the T
0
generation mode.
:FSYNc
DIS / T0 / CHA /
CHB / etc.
Sets the pulse source for the front panel sync output.
:RSYNc
DIS / T0 / CHA /
CHB / etc.
Sets the pulse source for the rear panel sync ouput.
:TRIGger
Subsystem. Contains the commands to define the
Trigger function.
:MODe
<boolean value>
Sets Trigger Mode. Disable or TRIG (enable).
:TSOurce
FRONtpanel /
REARpanel
Switches the system trigger source between front or
rear panel inputs.
:EDGe
FALLing / RISing
Selects which edge (rising or falling) to use as the
trigger signal.
:LEVel
<numeric value>
Sets the Trigger Threshold. Value is in volts, with a
range of .20 to 15 Volts.
:FENable
<boolean value>
Enables/Disables the trigger pulse width filter.
:FILTer
<numeric value>
Sets the trigger minimum pulse width value.
:COUNt
Subsystem. Contains commands to manipulate the
system shot counter (T
0
pulses)
:STATe
<boolean value>
Enables/Disables the system shot counter.
:CLear
0
Clears the system shot counter.
:COUNt
Query Only. Returns the value of the system shot
counter.
:PRETest
Query Only: Returns the measured resistance value
for all channel with PreTest enabled in the form of
<n>:x.xx with each channel separated by a single
space (e.g. 1:1.99 2:2.00 4:2.01 if all channels except
channel 3 have PreTest Enabled)
:POSTtest
Query Only: Returns the measured resistance value
for all channel with PostTest enabled in the form of
<n>:xxx.xx with each channel separated by a single
space (e.g. 1:149.99 2:147.00 4:150.01 if all channels
except channel 3 have PostTest Enabled)