Example: Control of the IQTX with SCPI
The following is a simple example of how to control the IQTX instrument by using SCPI commands. See
the previous section for specific details and extra parameters that the listed commands accept.
After any command, it is recommended to query the
*ESR?
command. This will allow debugging of
unreceived or incorrect commands that were sent to the product.
Identifying the IQTX instrument and installed modules:
1.
Query to confirm the correct instrument is setup
:*IDN?
-> Quantifi Photonics Ltd,IQTX-1102-FA-EPIQ,QP-998877,HW0.00.01SW3.01.18
2.
Query the available instrument module configuration
:*OPT?
-> IQABC,LASER
3.
Query the identification information for a specific slot module
:
MODUle2:IDN?
-> Quantifi Photonics Ltd,LASER-1002-2-SC,QP-987654,HW0.01.02FW0.01.32
Configuring the Laser module on the IQTX instrument:
1.
Set the Laser output power to 10 dBm
:
SOURce2:CHANnel1:POWer 10 DBM
2.
Set the Laser frequency to 193.4145 THz (1550nm)
:
SOURce2:CHANnel1:FREQuency 193.4145 THZ
3.
Turn the Laser ON
:
OUTPut2:CHANnel1:STATE ON
The following section details the various methods that a user may send these commands to the
IQTX instrument via
SCPI
commands
Force calibrate of the IQABC:
1.
Force the recalibration of the auto-bias control system
:ABC1:CALibrate
2.
Query the completion status of the auto-bias control system calibration
:ABC1:CALibrate? COMpleted
-> TRUE
Querying the status of the IQABC module:
1.
Query whether the XI bias is locked in the optimal position
:ABC1:XI:LOCK?
-> 1