Coherent Modulation Transmitter | IQTX
Quantifi Photonics Ltd.
Version
1.04
53
Command
:ABC:<XI|XQ|XP|YI|YQ|YP>:HISTory?
Syntax
:ABC:<XI|XQ|XP|YI|YQ|YP>:HISTory?<wsp>[<start_time>,<end_time><BIAS|E
RROR|FULL]
Description
Query the error history timeframe and reason
Parameters <wsp>: The start and end time of the error history
BIAS: The bias voltage value and time when the error occurred
ERROR: The error number and time when the error occurred
FULL: Returns all of the above parameters in a comma separated string
Response
<wsp>: <starttime>,<endtime>
BIAS: <timestamp[1]+startTime>:<vbias[1]>
ERROR: <timestamp[1]+startTime>:<error[1]>
FULL: <timestamp[1]+startTime>:<vbias[1]>,<error[1]>
Example
:ABC1:XI:HISTory? FULL -> 41000203.456:2.54,4.55
Command
:ABC:<XI|XQ|XP|YI|YQ|YP>:SCAN?
Syntax
:ABC:<XI|XQ|XP|YI|YQ|YP>:SCAN?<wsp>[POWER|ERROR|PHASE|FULL|UNIT]
Description
Query the ABC modulator scan results
Parameters POWER: The measured power values at the time of the scan for the current bias value
ERROR: The error numbers at the time of the scan for the current bias value
PHASE: The phase values at the time of the scan for the current bias value
FULL: Returns all of the above parameters in a comma separated string
UNIT: Get the scan measurement unit
Response
POWER: <number of points N>:<dcbias x-axis for all>:<power1,..,powerN>
ERROR: <number of points N>:<dcbias x-axis for all>:<error1,..,errorN>
PHASE: <number of points N>:<dcbias x-axis for all>:<phase1,...,phaseN>
FULL: <number of points N>:<dcbias x-axis for
all>:<power1,..,powerN>:<error1,..,errorN>:<phase1,...,phaseN>
UNIT: The scan measurement unit
Example
:ABC1:XI:SCAN? POWER -> 32:-14.000000,-13.125000,-12.250000,-
11.375000,-10.500000,-9.625000,-8.750000,-7.875000,-7.000000,-
6.125000,-5.250000,-4.375000,-3.500000,-2.625000,-1.750000,-
0.875000,0.000000,0.875000,1.750000,2.625000,3.500000,4.375000,5.2500
00,6.125000,7.000000,7.875000,8.750000,9.625000,10.500000,11.375000,1
2.250000,13.125000:0.000987,0.000926,0.000987,0.001094,0.000865,0.001
048,0.000865,0.000926,0.000835,0.000987,0.000774,0.001018,0.000850,0.
001109,0.000896,0.000896,0.000957,0.000926,0.000822,0.001021,0.143313
,0.143236,0.143236,0.143236,0.143267,0.143145,0.143221,0.143236,0.143
175,0.143099,0.143221,0.143130
Command
:ABC:<XI|XQ|XP|YI|YQ|YP>:GAIN
Syntax
:ABC:<XI|XQ|XP|YI|YQ|YP>:GAIN<wsp><value|MIN|MAX|DEF>
Description
Set the gain VPi percentage
Parameters Value: Set the gain VPi percentage value of the modulation channel
MIN: Set the minimum gain VPi percentage value of the modulation channel
MAX: Set the maximum gain VPi percentage value of the modulation channel
DEF: Set the default gain VPi percentage value of the modulation channel
Response
No response
Example
:ABC1:XI:GAIN SET 20.75