Specifications (Continued)
Safety Features
Ground
Continuity Test:
Programmable: 0.1
Ω
to 5.0
Ω
±
0.2
Ω
Accuracy, or OFF
Ground Fault
Interrupt (GFI):
Shutdown of current imbalance when I > 0.5mA
±
0.25mA, or OFF
In-Rush Current:
DC Mode: Set detection limit: 0.5uA – 5mA in 0.0001mA increments
The programmable range for In-Rush current is dependent on the
programmed High Limit:
Range: High
Limit:
In-Rush:
1 0.1
µ
A – 300
µ
A 0.5
µ
A – 300
µ
A
2
0.301mA – 3.000mA
5
µ
A – 3.000mA
3
3.01mA – 5.0mA
50
µ
A – 5.0mA
Fast Output Cutoff:
HV output voltage terminated <0.4mS after NG (Fail) result
Fast Discharge:
<0.2s (Typical) Discharge of DUT upon termination of HV.
ARC Detection:
Detection Current:
Range: 1mA – 20mA AC and 5mA DC
Pulse Width:
Minimum: 10
µ
s
General Features
Time:
Test*:
0.1sec – 999 sec, Continuous
Ramp:
0.1sec – 999 sec, OFF
Dwell:
0.1sec – 999 sec, OFF (DC & IR Mode only)
Fall:
0.1sec – 999 sec, OFF
* Test Time is limited
≤
60seconds when the voltage and high current limit
>
100VA.
* Test Time for IR is 0.3sec – 999sec, Continuous
Limits:
HI/LO programmable during Test Time
LO can be set to OFF during Hipot Test
HI can be set to OFF during IR Test
Indication:
Pass/fail LEDs, audible alarm
Remote Control:
Inputs:
START, STOP
Characteristics: 24V active low, Pulse width
≅
20ms
Outputs:
PASS, FAIL, UNDER TEST
Characteristics: Dry contact relay, Closed if true
115V,
<100mA
Connector: 9 pin male D-series & Terminal Strip