Operation
Page 51 of 85
OFFSET Function
There is no offset in the IR or PA modes. Using the pre-programmed example from paragraphs
2.4-2.8 of the GC, AC, DC, IR & PA 5-step test the resultant error and correction is illustrated
herein.
After PASS Screen, press [STOP] and screen reverts to STAND BY display.
Continue on Next Page.
GET
OFFSET
To GET OFFSET values
[F4]
To return to STAND BY status
To enter OFFSET function.
START
MORE..
OFFSET
RMT
ERR
OFST
LOCK
STAND BY
F1
F4
F3
F2
-------------
OUTPUT
1
2.750kV
15.00mA
3
2
AC
DC
IR
2.500kV
0.750kV
1.5M
Ω
2.999mA
-------------
-------------
4
5
GC
PA
M01
MEASURE
RESULT
0.1A
CHANGE-DUT
1.2
Ω
-------------
-------------
CANCEL
PRESS TEST KEY TO GET OFFSET..
RMT
ERR
OFST
LOCK
F1
F4
F3
F2
1.
2.
PLEASE OPEN THE HV
OUTPUT TERMINAL.
NO OFFSET IN IR MODE.
MORE..
PRESS TEST KEY TO CONTINUE..
RMT
ERR
OFST
LOCK
F1
F4
F3
F2
1.
2.
PAUSE: CHANGE-DUT
UNDER TEST SIGNAL: OFF
M01
5/5
PA
U.T.S = OFF
MORE..
RMT
ERR
OFST
LOCK
F1
F4
F3
F2
1.
2.
PAUSE: CHANGE-DUT
UNDER TEST SIGNAL: OFF
M01
5/5
PA
PASS
PASS
OFFSET
VALUES
MORE..
RMT
ERR
OFST
LOCK
U.T.S = OFF
F1
F4
F3
F2
PASS
OUTPUT
1
2.753kV
0.07mA
3
2
AC
DC
IR
2.499kV
0.750kV
1.5M
Ω
0.002mA
PASS
PASS
4
5
GC
PA
M01
MEASURE
RESULT
0.1A
CHANGE-DUT
0.1
Ω
SKIPPED
"STAND BY"
Display
with
Memory ID
PROGRAM
PRESET
MORE..
MENU
RMT
ERR
OFST
LOCK
STAND BY
0.1A
0.5s
1.2
Ω
F1
F4
F3
F2
LOW
GC
STEP 1/5
To access OFFSET function.
0.3
Ω
M01
:
START