3.3
S07 Power Entry Adapter Cable
The S07 Power Entry Adapter Cable is a 3-wire AC inlet receptacle for precise testing of corded
products. The S07 cable is connected to the Sentry Plus Series instrument via a two-lead set.
1.
Remove DUT’s power cord form its AC inlet module.
2.
Plug S07 Power entry adapter into DUT’s AC inlet module.
3.
Connect the white custom banana plug to the Sentry Plus OUTPUT terminal. Connect
black banana plug with retaining bracket to the Sentry Plus RTN/LOW terminal.
4.
For GC test, connect GC Lead (700100) to rear Sentry Cont Check terminal and metal on
case of DUT.
Figure 3-7 illustrates this connection of the S07 cable to a Sentry 30 Plus instrument.
Black
Banana
Plug with
Retaining
Bracket
to
RTN/LOW
Custom
White
Banana
Plug
to
OUTPUT
S07
Power Entry
Adapter
DUT
(Monitor)
Plug S07 Adapter into
DUT's AC inlet module
Remove DUT's power
cord
from AC inlet module
F1
RTN/LOW
F2
OUTPUT
DANGER
PASS
FAIL
F1
F2
F3
F4
START
STOP
Sentry 30 Plus AC/DC/IR Hipot Tester
0
1
Q
uadTech
!
CAUTION
F4
F3
Max 5kVAC
6kVDC
CAL
UPDATE
1
2
3
HV OUTPUT
!
HIGH VOLTAGE
Max 5kVAC/6kVDC
RTN/LOW
!
CONT. CHECK OPTION
VOLTAGE SELECTOR
POLLUTION DEGREE 2
INSTALLATION CATEGORY II
WARNING: FOR CONTINUED PROTECTION
AGAINST FIRE HAZARD, REPLACE ONLY WITH
THE SAME TYPE AND RATING OF FUSE AS
SPECIFIED FOR THE LINE VOLTAGE BEING
UTILIZED.
CAUTION: NO OPERATOR SERVICEABLE PARTS
INSIDE. REFER SERVICING TO QUALIFIED
PERSONNEL.
90V - 110V ~
216V - 250V ~
198V - 242V ~
108V - 132V ~
3.15AT
250V
1.6AT
250V
INTER
LOCK
COM
START
RESET
6
9
1
5
100V/120V/220V/240V~
50/60Hz 300W MAX
MODEL NO.
SERIAL NO.
FAN 50
o
C AUTO ON/OFF
Ground Cont
inui
ty
Connec
ti
on
REAR Sentry Plus Instrument
Hipot
Connection
4
Figure 3-7: S07 Power Entry Adapter Cable