Page 70 of 85
Interface
3.2
G16 International Power Strip
The Sentry Plus Series instruments can be connected to the G16 International Power Strip as
illustrated in Figure 3-6 for safety testing of many European corded products.
*Australia *United
Kingdom
*Denmark
*North
America
*Norway
*Finland
*Sweden
*Germany
*Netherlands
*Austria
*Switzerland
*Italy
1.
Connect the three G-16 ground connectors to the Sentry Plus RTN/LOW terminal.
2.
Note: there is a second RTN/LOW terminal on the rear of the Sentry instrument.
3.
Connect the white banana plug to the Sentry Plus OUTPUT terminal.
4.
Plug DUT’s AC power cord into corresponding module on G16 power strip.
Figure 3-6: G16 International Power Strip Connection to Sentry 30 Plus
G16 International Power Strip
Black
Banana
Plug
to
RTN
Green
Lug
to
RTN
Green/
Yellow
Lug
to
RTN
or RTN
on Rear
Panel
White
Custom
Banana
Plug
to
HV
OUTPUT
F1
RTN/LOW
F2
OUTPUT
DANGER
PASS
FAIL
F1
F2
F3
F4
START
STOP
Sentry 30 Plus AC/DC/IR Hipot Tester
0
1
Q
uadTech
!
CAUTION
F4
F3
Max 5kVAC
6kVDC
CAL
UPDATE
QuadTech
AUSTRALIA
UNITED KINGDOM
DENMARK
NORTH AMERICA
NORWAY FINLAND
SWEDEN
SWITZERLAND ITALY
GERMANY
AUSTRIA NETHERLANDS
Sentry 30 Plus Hipot Tester