Operation
Page 37 of 85
2.5
Programming an AC Hipot Test
This test is applicable to the Sentry 10, 20 and 30 Plus instruments. With the instrument in
‘stand-by’ status, press [F1] = PROGRAM. If continuing the 5-step example from ¶2.4, you are
already in program mode and on the AC test mode page. Follow the
green
arrows (
!
) on the
right side of this diagram to program the individual AC hipot test parameters.
Continue on next page.
INC.
DEC.
EXIT
NEXT
RMT
ERR
OFST
LOCK
0.05 - 5kV
F1
F4
F3
F2
FALL
RAMP
ARC
LOW
AC
STEP 2/2
VOLT
HIGH
TIME
Select
AC
Test Mode
Set
Test Voltage
Set
High Current
Limit
To set high current limit:
0.001 - 20mA in .001mA increments
To move highlighted box to HIGH.
To move highlighted box to TIME.
INC.
DEC.
EXIT
NEXT
RMT
ERR
OFST
LOCK
0.001 - 20mA
F1
F4
F3
F2
FALL
RAMP
ARC
LOW
AC
STEP 2/2
VOLT
HIGH
TIME
INC.
DEC.
EXIT
ENTER
RMT
ERR
OFST
LOCK
0, 0.1 - 999s
F1
F4
F3
F2
FALL
RAMP
ARC
LOW
AC
STEP 2/2
VOLT
HIGH
TIME
INC.
DEC.
EXIT
NEXT
RMT
ERR
OFST
LOCK
0 - 20mA 0 = OFF
F1
F4
F3
F2
FALL
RAMP
ARC
LOW
AC
STEP 2/2
VOLT
HIGH
TIME
To set test time:
0, 0.1 - 999s in 0.1s increments
To move highlighted box to LOW.
Set
Test Time
To move highlighted box to ARC.
To set low current limit:
0 - 20mA in 0.01mA increments
To set test voltage:
0.05 - 5kV in .001kV increments
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
To move highlighted box to VOLT
Set MODE = AC
UP
DOWN
EXIT
NEXT
RMT
ERR
OFST
LOCK
Select Mode
F1
F4
F3
F2
FALL
RAMP
ARC
LOW
AC
STEP 2/2
VOLT
HIGH
TIME
OFF
OFF
OFF
OFF
[F1] = New
Program
Step 2 = AC
To move highlighted box around
display to select parameter to change
To exit programming mode.
To delete or insert a test step.
To change value in highlighted box.
FIRST
MORE..
EXIT
NEXT
RMT
ERR
OFST
LOCK
Voltage is 0
F1
F4
F3
F2
FALL
RAMP
ARC
LOW
AC
STEP 2/2
VOLT
HIGH
TIME
OFF
OFF
OFF
OFF
0.000kV
:
:
:
:
:
:
:
0.500mA
3.0s
:
:
:
:
:
:
:
3.0s
0.500mA
0.000kV
0.000kV
:
:
:
0.500mA
3.0s
:
:
:
:
:
:
:
:
:
:
:
2.750kV
0.500mA
3.0s
:
:
:
3.0s
15.00mA
2.750kV
:
:
:
:
Set
Low Current
Limit
2.750kV
:
:
:
10.0s
15.00mA
:
:
:
:
Test time = 0
then continuous
voltage until
STOP
is pressed.