![Qiagen EZ1 Advanced XL User Manual Download Page 90](http://html2.mh-extra.com/html/qiagen/ez1-advanced-xl/ez1-advanced-xl_user-manual_3168624090.webp)
EZ1 Advanced XL User Manual 11/2017
90
Appendix A
Technical data
QIAGEN reserves the right to change specifications at any time.
Operating conditions
Power
100–120 V AC, 50/60 Hz, 600 VA (North America and Japan)
200–240 V AC, 50/60 Hz, 600 VA (Europe)
Mains supply voltage fluctuations are not to exceed 10% of nominal
supply voltages.
Fuses
6.3 A (250 V) time-delay fuse (for 100–120 V AC)
3.15 A (250 V) time-delay fuse (for 200–240 V AC)
WARNING
Electrical hazard
Never install a fuse different from that specified in the user manual.
Overvoltage category II
Air temperature
15–30ºC (59–86ºF)
Relative humidity
15–75% (noncondensing)
Altitude
Up to 2000 m (6500 ft.)
Place of operation
For indoor use only
Pollution degree
2
Environmental class
3K2 (IEC 60721-3-3)
3M2 (IEC 60721-3-3)
Transportation conditions
Air temperature
–25ºC to 60ºC (–13ºF to 140ºF) in manufacturer’s package
Relative humidity
Minimum of 15% to maximum of 75% (noncondensing)
Environmental class
2K2 (IEC 60721-3-2)
2M2 (IEC 60721-3-2)