![Qiagen EZ1 Advanced XL User Manual Download Page 69](http://html2.mh-extra.com/html/qiagen/ez1-advanced-xl/ez1-advanced-xl_user-manual_3168624069.webp)
EZ1 Advanced XL User Manual 11/2017
69
The following screen appears.
SETUP: REMINDER PM
Interval
Up, Dn, ENT, ESC
Press the
up arrow
or the
down arrow
to change the
Interval
field to
1/2 year
(semiannual) or
1 year
(annual).
After setting the new interval, store the setting by pressing
ENT
. Alternatively, press
ESC
to leave the
current interval unchanged.
5.11
Using the bar code reader
The bar code reader is preset to read the following types of bar codes.
2/5 interleaved
EAN Family
Code 39
Code 128
To read a bar code, press the button located on the inner side of the handle of the bar code reader.
A red light appears. Hold the bar code reader in front of the bar code at a distance of
approximately 20 mm. A beep will sound to confirm that the bar code has been read. The bar
code will be displayed on the screen.
Length of barcode label: 15–65 mm
Minimum tube diameter: 9 mm
Maximum resolution of bar code: 0.1 mm
5.12
Using the external printer
Printers sold in the open market can be connected to the instrument.
The EZ1 Advanced XL allows you to confirm that a report was printed correctly before the report
file is deleted. The following screen appears.
SEND REPORT
Print out o.k ?
1: o.k.
2: not o.k.
ESC: Back