Multi-Pointer
Service Manual
Page 22 of 26
State 7
State 8
Fig. 32
4. When the front, left and right laser point error is as shown in the Fig. 32, you need
to adjust the power adjusting screw to calibrate (see Fig. 33). When the instrument
is in the state 7, you can loose the screw 7 and screw down 8. When the instrument
is in the state 8, you can loose the screw 8 and screw down 7.
Fig. 33