9.7.2
Check zeros – zero using N
2
START
Switch on the
instrument and
complete the
instrument warm up.
Main Gas Read
Screen select the
‘Menu’ key
Select key 2
‘Gas Check’
Select key 1
‘Check Zeros’
Select
soft-key
‘Select Gas’
Wait for the readings to
stabilise. Wait for the
analyser to zero for
approx. 1 min.
Message will display
‘Please wait for timer
to reach zero’
When complete a
message will
display ‘Check
complete please
confirm results’
READ STATUS
FIELD
Select soft-key
‘Confirm’
READ STATUS
FIELD
Zero complete?
Zero complete?
Result saved
END
Select
soft-key
‘Check’
NO
Select
soft-key
‘Exit’
Zero Complete
YES
Select
soft-key
‘Exit’
YES
NO
Select ‘Zero using N2’
Select soft-key
‘Retry’
(retry zero again)
Select soft-key
‘Zero’
(set to zero)
OR
Contact Geotech
technical support team
for advice
YES
Operation Settings
screen
Flow N2
Connect the gas
regulator supplied to
the white port on the
analyser
Summary of Contents for BIOGAS 5000
Page 44: ...5K MNL BIOGAS5000 BIOGAS 5000 gas analyzer Page 44 8 9 Taking gas and flow measurement...
Page 45: ...BIOGAS 5000 Gas Analyzer 5K MNL BIOGAS5000 Page 45...
Page 53: ...BIOGAS 5000 Gas Analyzer 5K MNL BIOGAS5000 Page 53...
Page 61: ...BIOGAS 5000 Gas Analyzer 5K MNL BIOGAS5000 Page 61...
Page 82: ...5K MNL BIOGAS5000 BIOGAS 5000 gas analyzer Page 82 Dr Roger Riley...
Page 103: ...BIOGAS 5000 Gas Analyzer 5K MNL BIOGAS5000 Page 103 Dr Roger Riley...