CARING FOR YOUR CAMERA
CHAPTER 5
The
Rolera EM-C
2
camera requires no regular maintenance except occasional external cleaning of the
cooling window (the glass window that forms part of the cooling chamber and sits between the
camera sensor
and the microscope or lens).
To clean the Cooling window:
CAUTION — The camera’s Cooling windows, CCD sensor, and circuits are
sensitive to static discharge. Ensure you are using a static strap or are
completely grounded at all times to release any static energy before you
clean the cooling window.
•
Use clean
forced air (available at stores that sell cameras and computer cleaning supplies)
to dust the cooling window (recommended).
•
If the image still appears dirty, gently wipe the face of the cooling window with a small
amount of optical grade isopropyl alchohol and lens paper.
•
Apply forced air again to remove any loose particles.
12
Summary of Contents for Rolera EM-C2
Page 1: ......