Motor & Auger Assembly
1.
Place the motor mount plate (4) onto the bottom of the servo motor (16) and
assemble using three machine screws (12) tightening with a 1.5 mm Allen key.
2.
Thread the lock knobs (5) into the motor mount plate (4).
3.
Slide the flexible coupling (17) onto the shaft of the servo motor (16) then use the
0.9 mm Allen key to tighten the two set screws on the flexible coupling to secure
it in place.
Note: Be sure one of the set screws are secured to the flat of the servo motor shaft.
4.
Mount the o-ring (18) onto the auger (14) by sliding it over the top, non-threaded
portion of the auger.
Note: Do not slide the o-ring over the threads of the auger.
5.
Insert the top, non threaded portion of the auger (14) into the coupling (17) and
secure it in place by tightening the two set screws using a 0.9 mm Allen key.
Note: be sure one of the set screws are secured to the flat of the servo motor shaft.
Fluid Body Assembly
1.
Using a press and press tool (114-5375), place the press tool into the male side of
the luer adapter (15), align the inlet hole of the luer adapter to the inlet hole of the
fluid body (1) and press the two pieces together until the luer is inserted all the
way to the hard stop.
Note: After assembly, look through the inlet hole of the fluid body to verify the
holes are aligned properly.
2.
Thread the four standoffs (9) into the fluid body (1) and tighten using an
adjustable wrench.
3.
Place the syringe support (8) onto the standoffs (9) with the countersink away
from the standoffs and assemble using four machine screws (12) tightening with a
3/32” Allen key.
Note: The bracket should extend out on the same face as the port of the fluid body
4.
Assemble the quick release assembly by first placing the release lock right (2)
component onto the fluid body (1) then inserting the spring (13) in place. Then
while holding the spring in place, assemble the quick disconnect cover (6) onto
the assembly and secure with the machine screws (11) using a 0.050” Allen Key
to tighten.
5.
Repeat step 4 on the left side of the valve with the release lock left (3), spring
(13), quick disconnect cover (6), and machine screws (11).
6.
Place the disconnect bar (7) in between the two release lock levers (2, 3) and
secure the assembly with the two machine screws (10) using a 1/16” Allen Key to
tighten.