OPERATING MANUAL FOR SELECT FUSION
7.4 REPLACING 0.2 µm POINT OF USE FILTER.
Fig-29
•
Fitting of a new filter can be
carried out without turning the unit
off.
•
Simply hold the body of the
capsule and un-screw clockwise.
•
Once unscrewed dispose of the
filter and remove the new one
from its packaging. Refer to
Section 9.4 for advice on
disposal.
•
Wind about 4 turns of PTFE tape
around the thread on the filter.
•
Screw the new filter anti-clockwise back into the dispense valve.
(NB: Be careful not to cross thread the filter)
•
The bleed nozzle should point out to the right hand side of the unit.
•
To bleed the filter of air, open the bleed screw a few turns, remove
the lower bell-housing cap and dispense approx 500 mls of water,
tapping the filter gently.
•
When all of the air has gone close the bleed screw.
Iss: F 17.03.08