OPERATING MANUAL FOR SELECT FUSION
1.4
TREATED WATER SPECIFICATION
Parameter
High Purity
Dispense
Tank
BS EN ISO 3696
Grade 1+
Grade II
ASTM D1193-99e1
Type-1
Type III
Resistivity @ 25
o
C 15-18.2M
Ω
-cm 1-15
M
Ω
-cm
PH 6-8 6-8
TOC (ppb as C)
<30
<50
Bacteria <1
cfu/ml <99%rejection
Particles
0.2 um Free
N/a
Organics <0.001AU@254nm
N/a
Residual Solids (ppm) <0.1
<0.5
1.5 WATER
SERVICES
CONNECTIONS
Description
Connection
Drain
8mm Pushfit
Mains Water In
8mm Pushfit
Overflow
12mm
Pushfit
1.6
WEIGHTS & DIMENSIONS
Unit
Weight kg
Height
(mm)
Width (mm)
Depth (mm)
Fusion
40/80/160
22 (Dry)
42 (Working)
630 440 548
Fusion
40/80/160 with
boost pump
24 (Dry)
44 (Working)
630 440 548
1.7 ENVIRONMENTAL
Room storage and operating
temperature range
5 to 40
o
C (41-104
o
F)
Relative Humidity
30 to 80%
Max Altitude
2000m
Transport and Storage
temperature (limited by RO
membranes)
-5 to 85
o
C (with frost protection liquid
– 40 to 85
o
C)
RFI/EMI Radiation
The ‘EMC’ environment must be
within the limits to which the unit has
been tested, see section 1.7.
Care must be taken not to have
sources of RFI/EMI, which are liable
to cause electromagnetic disturbance
to the unit. If the unit is affected by
such disturbances, the sources
should be suppressed or relocated.
Iss: F 17.03.08
Summary of Contents for SELECT FUSION 160
Page 1: ...SELECT FUSION OPERATOR S MANUAL ISSUE F 17 03 08...
Page 40: ...OPERATING MANUAL FOR SELECT FUSION Iss F 17 03 08...
Page 41: ...OPERATING MANUAL FOR SELECT FUSION Iss F 17 03 08...
Page 42: ......
Page 63: ...OPERATING MANUAL FOR SELECT FUSION Iss F 17 03 08...
Page 71: ...OPERATING MANUAL FOR SELECT FUSION Iss F 17 03 08...
Page 76: ...OPERATING MANUAL FOR SELECT FUSION Iss F 17 03 08...
Page 85: ...OPERATING MANUAL FOR SELECT FUSION Iss F 17 03 08...
Page 86: ...OPERATING MANUAL FOR SELECT FUSION Iss F 17 03 08...
Page 87: ...OPERATING MANUAL FOR SELECT FUSION Iss F 17 03 08...
Page 88: ...OPERATING MANUAL FOR SELECT FUSION Iss F 17 03 08...