PureAire Monitoring Systems, Inc.
10
3: Installation
3.1 Site Requirements
The Universal Gas Detector enclosure should be mounted in an area free of vibration and electrical
noise or interference. If possible, avoid areas with high temperatures or condensing humidity.
WARNING: The Universal Gas Detector is not designed for installation in
hazardous areas.
3.2 Mounting
3.2.1 Transmitter Enclosure
The Universal Gas Detector is designed primarily for wall mounting and should be installed at a height
convenient for operation, maintenance, and viewing of the instrument display. The following is a
drawing of the mounting dimensions.
3.2.2
Universal Gas Detector
The transmitter and sensor should be installed in a location where gas leaks are likely to occur or
where released gases may accumulate. It should be mounted no closer than 12 inches above floor level.
Airflow within the monitored area, the characteristics of the gas (lighter or heavier than air), and the
position of workstations and personnel should all be considered in determining the most suitable
installation location.
8.
1
2
5
”
5.125”